104 Keysight N9010A EXA Service Guide
Instrument Messages
-200 to -299, Execution Errors
-200 to -299, Execution Errors
For -221 error messages, see the previous sections.
Note that Execution Errors are divided into subclasses:
-21x – Trigger errors
-22x – Parameter error
-23x – Data corrupt or stale (invalid data)
-24x Hardware error
-25x – Mass storage error
-26x – Expression data error
-27x – Macro error
-28x – Program error (a downloaded program-related execution error)
-29x – Memory use error
Err# Message Verbose/Correction Information
-200 Execution Error A program execution error has occurred. The exact problem cannot be
specifically identified.
-200 Execution error; Carrier
frequency outside device’s
transmit band
The entered channel/carrier frequency is not within the range of your
current mode setup selection of standard and device.
-200 Execution error; Invalid
GSM burst timing
A GSM-like burst was acquired, but its timing is not valid. Ensure the
correct Burst Type has been selected.
-200 Execution error; Invalid IP
address
The IP address supplied is either not valid or does not belong to a
compatible Signal Generator. Please check the IP address and
instrument connection and try again.
-200 Execution error; Invalid
Marker Trace.
Cannot place markers on the reference trace, because the reference
trace is currently turned off or has no data.
-200 Execution Error; No peak
found.
No signal peak was found within the defined parameters of the search.
-200 Execution error; No ranges
are defined. Activate a
range.
There are no active ranges in the range table. You will need to activate
at least one range.
-200 Execution Error;
Preselector centering
failed
Algorithm failed to center the preselector. This maybe caused by the
signal peak being too low in amplitude. Or it could be from excessive
CW input signal, alignment error, or hardware failure.
-200 Execution Error; Signal not
stable enough to track
The signal that you have selected to track is changing too much for the
function to track it properly.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals