Keysight N9010A EXA Service Guide 103
Instrument Messages
-221 Settings Conflict Errors
-221 Settings conflict; Tx Band
Spur measurement is not
defined for mobiles.
Only base station testing is available.
-221 Settings conflict; Zero Span
not available when Frequency
Scale Type = Log
Logarithmic scales cannot be used for time domain sweeps (0 Hz
span).
-221 Settings conflict; Zero Span
not available when Segmented
Sweep is on
Segmented Sweep uses multiple sweeps. It cannot combine
sweeps with 0 Hz frequency (time domain sweeps).
-221 Settings conflict; Zone is
not available when Segmented
Sweep is on
The Zone functionality cannot be used while using segmented
sweep.