472 Keysight N9010A EXA Service Guide
Assembly Replacement Procedures
Top Brace and Reference Bracket
Top Brace and Reference Bracket
Removal
1. Remove the instrument outer case. Refer to the “Instrument Outer Case”
removal procedure.
2. Refer to Figure 15-6. On analyzers with serial prefix < MY/SG/US5648,
both the Top Brace (1) and the Reference Bracket (2) will be present. On
analyzers with serial prefix MY/SG/US5648 only the Top Brace (1) will
be present.
3. To remove the Top Brace, use the T-10 driver to remove the eight screws
(3) (0515-0372) attaching the Top Brace to the chassis. Remove the six
flathead screws (4) (0515-1227) attaching the Top Brace to the A7
Midplane, A14 LO Synthesizer, and A15 Front End Controller assemblies.
Remove the four flathead screws (7) (0515-1227) securing the Top Brace
to either the Reference Bracket or the A16 Reference assembly.
4. To remove the Reference Bracket (2) (if present), use the T-10 driver to
remove the six screws (5) (including two at the rear panel) attaching the
Reference Bracket to the instrument.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals