Keysight N9010A EXA Service Guide 471
Assembly Replacement Procedures
Instrument Outer Case
Replacement
1. Disconnect the instrument from ac power.
2. Slide the instrument cover back onto the deck from the rear. The seam on
the cover should be on the bottom. Be sure the cover seats into the gasket
groove in the Front Frame Assembly.
3. Refer to Figure 15-4. Replace the strap handle plugs (5) in place and
attach using the T-20 driver to replace the four screws (two on each side)
(6). Torque to 21 inch pounds.
4. Place the bail handle into position on the front frame and using the T-20
driver, replace the four screws (two on each side) (1) that hold the bail
handle (2) to the front frame. Torque to 21 inch pounds.
5. Refer to Figure 15-5. Replace the four rear bumpers to the rear of the
instrument. Torque to 21 inch pounds.