Keysight N9010A EXA Service Guide 177
RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
Troubleshooting a High Band Problem
1. Reference Assembly Verification
Refer to Figure 4-22. Remove cable W19 from A9 Input Attenuator A (1).
Measure the 4.8 GHz calibrator signal on the cable end with a functioning
Spectrum Analyzer.
Figure 4-22 W19 Location