Keysight N9010A EXA Service Guide 91
Instrument Messages
-300 to -399, Device-Specific Errors
-300 to -399, Device-Specific Errors
Err# Message Verbose/Correction Information
-300 Device-specific error An instrument error occurred and the exact problem cannot be
specifically identified. Report this error to the nearest Keysight
Technologies sales or service office.
-310 System error; An internal system-type error has occurred. The exact problem cannot
be specifically identified. Report this error to the nearest Keysight
Technologies sales or service office.
-310 System error; A license
will soon expire;<feature
code> will expire in <time>
The indicated feature/software will expire in the specified time.
Contact Keysight Technologies to purchase continued use of this
functionality.
-310 System error; Error
transmitting a LAN event to
the network.
Communication with the network driver failed.
-310 System Error; Ext Source
not responding
Signal source at given IP address is not responding / IP does not
belong to a source. Check IP address and network connection.
-310 System error; Failed to
initialize the PTP clock to
current time.
Failure communicating with the DMC libraries’ PTP controller.
-310 System error; Failed to
instantiate the PTP
ordinary clock.
Failure in the starting up the DMC libraries’ PTP controller.
-310 System error; feature
<feature code> not licensed
The specified feature, for example “N9073A-TR2” is not licensed. The
license may have expired. You cannot use it until you get a license.
-310 System error; Feature
expired; <feature code>
The specified feature has expired. The license is no longer valid.
-310 System error; License
installation
failed;<feature code>
The license installation of the specified feature, for example
"N9073A-TR2", has failed. You should refer to the event log in the
control panel for more details.
-310 System error; License
removal failed; <feature
code>
The license removal of the specified feature, for example
“N9073A-TR2” has failed. You should refer to the event log in the
control panel for more details.
-310 System error; No license;
<feature code> will
terminate in <time>
The specified feature will stop working in the specified time due to the
license expiration You will be prompted to save results and exit.
-310 System error; The
configured PTP hardware
driver could not be
instantiated.
The PTP driver failed on initialization.

Table of Contents