92 Keysight N9010A EXA Service Guide
Instrument Messages
-300 to -399, Device-Specific Errors
-310 System error; The PTP
hardware driver reported a
configuration error.
Failure in the execution of the PTP driver. The most likely cause of this
error is a mismatch between versions of the PTP driver and the LXI
middleware.
-310 System error; The PTP
ordinary clock reported a
configuration error.
Failure in execution of the DMC libraries’ PTP controller.
-310 System error; The Trigger
alarm delayed LAN event was
not scheduled due to an
existing pending event.
Delayed LAN events cannot occur too close together (within 20 ms).
-310 System error; The Trigger
alarm delayed LAN event was
not scheduled due to a
conflict with an existing
scheduled alarm.
Delayed LAN events cannot occur too close to a scheduled Alarm
(within 20 ms).
-310 System error; The Trigger
alarm was not scheduled due
to a conflict with an
existing scheduled alarm.
Alarms cannot be scheduled to happen too close together (within 20
ms).
-311 Memory error There is a physical problem with the instrument memory, such as a
parity error.
-312 PUD memory lost Protected user data saved by the *PUD command has been lost.
-313 Calibration memory lost The nonvolatile calibration data used by the *CAL? command has been
lost.
-314 Save/recall memory lost The nonvolatile data saved by the *SAV? command has been lost.
-315 Configuration memory lost The nonvolatile configuration data saved by the instrument has been
lost.
-320 Storage fault; A problem was found while using data storage. The error is not an
indication of physical damage or failure of any mass storage element.
-321 Out of memory An internal operation needed more memory than was available. Report
this error to the nearest Keysight Technologies sales or service office.
-330 Self-test failed A self-test failure occurred. Report this error to the nearest Keysight
Technologies sales or service office.
-340 Calibration failed The instrument requires an Align All Now. Restore the alignment by
pressing System, Alignments, Align All Now.
-340 Calibration failed; no
signal detected; <port>
The calibration for the <I | I-bar | Q | Q-bar> port did not succeed
because the calibration signal was not detected. Check that the < I |
I-bar | Q | Q-bar > port is properly connected to the Cal Out port.
-350 Queue overflow An error occurred that did not get put in the error queue because the
queue was full.
-360 Communication error There was a problem with instrument remote communications. The
exact problem cannot be specifically identified.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals