Keysight N9010A EXA Service Guide 93
Instrument Messages
-300 to -399, Device-Specific Errors
-360 Communication error; SNS
data read failure.
Disconnect then reconnect
SNS
The Keysight Smart Noise Source connected to the MXA has failed to
be read by the application. Please disconnect and reconnect the SNS.
If this continues to fail, then the SNS may have had its EEPROM
corrupted or another hardware fault exists. Check SNS on another
instrument, NFA and ESA are also SNS compatible instruments. Check
the device is not an Keysight power sensor which uses the same cable
interface.
-361 Parity error in program
message
A parity bit was not correct when the data was received. For example,
on a parallel port.
-362 Framing error in program
message
A stop bit was not detected when data was received. For example, on a
remote bus port.
-363 Input buffer overrun A software or hardware input buffer on a port overflowed with data
because of improper or nonexistent pacing.
-365 Time out error There was a time-out problem in the instrument. The exact problem
cannot be specifically identified.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals