94 Keysight N9010A EXA Service Guide
Instrument Messages
-221 Settings Conflict Errors
-221 Settings Conflict Errors
This is one of the errors in the standard SCPI error range of -200 to -299. See
the table in section “-200 to -299, Execution Errors” on page 104 for the rest
of those errors.
The <subtext> part of a Settings Conflict error should be worded so that the
text is: “function1” is not whatever/with/while/when “function2”. This makes
them easier to find - alphabetically, to avoid duplicates.
The entire message displays in the error history as “-221, Settings conflict;
<subtext>”
For example, -221.0076 displays as:
-221, Settings conflict; Invalid trace number
Err# Message Verbose/Correction Information
-221 Channel BW is not available
in Zero Span
The Channel BW function is not available in zero span.
-221 De-emphasis only available
in FM
The de-emphasis function is only available if FM demod is
selected.
-221 Setting conflict; <trigger
source> trigger is not
available while input is
<input port>
The trigger source (Video, RF Burst, I/Q Mag, etc.) is not available
with the current input port (RF, IQ, etc.)
-221 Setting conflict;
Differential setting
determined by probe type
A probe is connected that has a built in Differential setting. The
setting cannot be changed manually.
-221 Setting conflict; Input Z
unavailable when probe
sensed
A probe is connected and the Input Z is set based on the probe
type. It cannot be changed manually.
-221 Settings conflict; A legal command was received but it could not be executed due to
the current device state.
-221 Settings conflict; *.CSV
file format is not available
in this measurement.
You cannot load or save base instrument traces, as this is not
supported by the Log Plot measurement.
-221 Settings conflict; <Q Param>
cannot be changed when Q
same as I
When the “Q Same as I” parameter is set to Yes, the I parameter
value is copied to <Q Param> and the <Q Param> value cannot be
changed. Set Q Same as I to No to enable explicit control of the <Q
Param> value.
-221 Settings conflict; A Valid
User Cal is required.
Optimize aborted
Optimize Preselector can only be performed if a valid user cal
exists and is applied to current results. Perform a user cal first or
apply existing cal.