206 Keysight N9010A EXA Service Guide
RF Section Troubleshooting (Millimeter-Wave Analyzers)
RF Section Description
2. RF input frequencies from 3.6 GHz to 17.1 GHz go through the high band
path. Refer to the RF Highband Path #1 Block Diagram (Millimeter-Wave
Analyzers) in Chapter 12 for details.
The input signal level can be optimized by Input Attenuator A and/or Input
Attenuator B. The Low Band Switch assembly routes the signal through
the A12 YTF Preselector. If one of the optional high band preamps is
licensed, the signal can be routed through the preamp. The A15 Front End
Controller provides switching signals.
The Low Band Switch assembly performs the following tasks:
Provides a limiter that offers added protection for the mixer and
optional preamp.
Contains the high band preamp.
Routes the input signal to either the low band input of the A13 Front
End assembly or to the A12 YTF Preselector. If Option MPB is
installed, the signal is routed through a transfer switch, SW4.
The high band input switch in the Front End Assembly routes the RF input
signal to Mixer #2, where it mixes with the LO and provides a 1st
intermediate frequency of 322.5 MHz. The 322.5 MHz IF output is at
A13J7. The control voltages and biasing are provided by the A15 Front End
Control Assembly. The LO is doubled for tuned frequencies > 8.4 GHz in
this path.
3. RF input frequencies from 17.1 GHz to 34.5 GHz go through the high band
path. Refer to the RF Highband Path #2 Block Diagram (Millimeter-Wave
Analyzers) in Chapter 12, “Block Diagrams.” for details.
The signal path for 17.1 GHz to 34.5 GHz operation is almost the same as
the 3.6 GHz to 17.1 GHz path. The only difference is that Mixer #3 is used
in this path. The LO is doubled for all tuned frequencies in this path.
4. RF input frequencies from 34.5 GHz to 44 GHz go through the high band
path. Refer to the RF Highband Path #3 Block Diagram (Millimeter-Wave
Analyzers) in Chapter 12, “Block Diagrams.” for details.
The signal path for 34.5 GHz to 44 GHz operation is almost the same as
the 3.6 GHz to 17.1 GHz path. The only difference is that Mixer #4 is used
in this path. The LO is doubled for all tuned frequencies in this path.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals