Keysight N9010A EXA Service Guide 39
Overview
Service Equipment You Will Need
Step Attenuator Loss Characterization
The step attenuator combination should have each attenuator setting
characterized by a metrology lab at 50 MHz. The following tables show which
sections of the 10 dB and 1 dB step attenuators are utilized for each attenuator
setting. The tables also list the Allowable Uncertainty for each attenuator
setting. The interconnect cable should NEVER be disconnected once the loss
characterization is performed.
Table 1-2 1 dB Step Attenuator
Nominal Attenuation (dB) Attenuator Section Allowable Uncertainty (dB)
#1
(1 dB)
#2
(2 dB)
#3
(4 dB)
#4
(4 dB)
0 Off Off Off Off 0 (Reference)
1 On Off Off Off < 0.005
2 Off On Off Off < 0.005
3 On On Off Off < 0.005
4 Off Off
On
a
Off < 0.005
5OnOff
On Off < 0.005
6OffOn
On Off < 0.005
7OnOn
On Off < 0.005
8 Off Off On On < 0.005
9 On Off On On < 0.005
a. TME requires that Section 3 be characterized.
Table 1-3 10 dB Step Attenuator
Nominal Attenuation (dB) Attenuator Section Allowable Uncertainty (dB)
#1
(10 dB)
#2
(20 dB)
#3
(40 dB)
#4
(40 dB)
0 Off Off Off Off 0 (Reference)
10 On Off Off Off < 0.010
20 Off On Off Off < 0.015
30 On On Off Off < 0.020
40 Off Off
On
a
Off < 0.025
50 On Off
On Off < 0.030

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals