Keysight N9010A EXA Service Guide 621
Post-Repair Procedures
Post-Repair Procedures
64-Bit Processors
The N9010A-PC6 and N9010A-PC7 Processors are 64-bit processors; all
earlier processors are 32-bit processors.
Default BIOS Settings
1. Connect an external USB keyboard to one of the front panel USB ports.
2. Turn on instrument power.
3. Confirm Keysight Technologies splash screen comes up within a few
seconds.
4. Press the key on the keyboard specified on the initial boot screen to enter
the setup utility.
5. Navigate to the Exit screen and select "Restore Defaults" or "Load Setup
Defaults", whichever is present.
6. Once the defaults have been loaded select "Save Changes" and "Exit".