12 Maintenance and Troubleshooting
d. Adjust the regulator pressure control knob until the Purge fill or Cycle fill pressure,
whichever is higher, is shown on the regulator display.
e. Allow gas to flow for 30 seconds.
f. Close the fill, expand, and vent valves.
g. Turn the regulator control knob clockwise to increase pressure by 2 psi (15 kPa).
12 - 14
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) — Jan 2021