Description
The range in psig is 0 to 500.
The range in kPag is 0 to 3450.
Number of Cycles
Key in the number of cycles to be performed. A cycle is a series of
functions which produce a single volume measurement.
The range is 1 to 999. For calibration, 10 is sufficient.
Cycle fill pressure
Key in the fill pressure. For most applications, the default of 19.500
psig is adequate.
The range in psig is 0 to 19.850.
The range in kPag is 0 to 136.86.
High Pressure sample chamber:
The range in psig is 0 to 500.
The range in kPag is 0 to 3450.
P1, P2 end by?
[selection]
Choose the manner in which to end the pressure measurement.
Press CHOICE to make a selection.
The available options are Equilibrate and Fixed interval.
Equilibration Rate: (Does
not display for Fixed
Intervals)
Select Equilibrate at the P1, P2 end by? prompt. Pressure
measurement will end when the entered rate is obtained. Key in the
equilibration rate.
A high rate will produce faster results, but results may not be as
precise as desired. The lowest rates may cause errors when some
materials (such as those with appreciable vapor pressures or
organics) are analyzed.
The default setting is 0.0050 psig/min.
The range in psig/min is 0.0001 to 9.0000.
The range in kPag/min is 0.0007 to 62.0500.
High Pressure sample chamber:
The range in psig is 0.001 to 99 minimum (the default is 0.05)
Interval Time: (Does not
display for Equilibrate)
Select the Fixed interval at the P1, P2 end by? prompt. Pressure
measurement ends when the specified time is reached. Key in the
interval time.
The range is 10 to 99999 seconds.
Analysis Parameters
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) Jan 2021
2 - 11

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ACCUPYC II 1345 and is the answer not in the manual?

Micromeritics ACCUPYC II 1345 Specifications

General IconGeneral
BrandMicromeritics
ModelACCUPYC II 1345
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals