2 Data Entry with Keypad
Display Text
Description
Use Run Precision?
Allows early termination of the analysis when certain criteria are met.
n Yes. The analysis terminates after five consecutive cycles are
within the specified tolerance. Always request a large number (50
to 99) of runs. The number of runs is determined through the Num-
ber of Cycles option. With a small number of runs, the analysis
stops when the selected number is reached, even if the specified
tolerance has not been met.
n No. Run Precision is not used.
Percent Full Scale:
This prompt only displays if Yes is selected for Run Precision and
more than five cycles have been requested.
Enter the run precision volume tolerance which is expressed as a
percentage of nominal cell volume (sample capacity).
The range is 0.01% to 50.00%.
2 - 12
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) Jan 2021