135
Everyday
All metal parts
(cover, clamps
and reactor)
Spray a bit of
silicon grease on
paper and pass it
on all metal parts
Before
long
operational
brakes
Internal high
pressure circuit
Run a program with 200 ml of DI water with 5 ml of
conc. HNO
3
in the big TFM vessel without rack and vials.
MW program:
Step
Time
T1
T2
P
Power
1
00:15:00
200°C
60°C
100 bar
1500W
2
00:05:00
200°C
60°C
100 bar
1500W
Every
months
TFM Vessel
Cleaning procedure to
remove NOx
Put the big PTFE vessel
in an electrical oven at

Table of Contents