Maintenance by Milestone customer service must be performed once a year or after a
maximum of 600 runs.
The number of cycles varies according to the application adopted. Ask for more information
from the service technician.
NOx decontamination of the main PTFE vessel
To remove the NOx gases from the PTFE vessel, run the following procedure:
• Clean the vessel with a paper moistened of water
• Dry it very well and put it into a conventional electric (not microwave) oven overnight (or at least
4 hours) at 160°C
Be aware that, during the cleaning in oven, a lot of NOx fumes could be released from the TFM
vessel.
Cleaning before long operational Breaks
Before leaving the instrument off for more than 5 days
Without rack inside, put 200ml of deionized water with 5 ml of HNO3 conc. In the PTFE vessel and run
the following heating program:
Load pressure: 40 bar
Open reactor and remove the water and dry all parts.
In order to clean the complete system before long operational break (≥5 days off), the system
have to be cleaned running the following procedure.