42
Pneumatic booster connected to Nitrogen generator
Gas
source
Standard
pressure
regulator
Connection
with
booster
Air
compressor
Pneumatic
Booster
ultraWAVE
3 unit
Nitrogen
generator
Standard
pressure
regulator (0-10
bar)
Pneumatic
booster requires
4-10bar pressure
Polymer
tube: 2m
length
6mm (ID)
8mm (OD)
Air compressor
with pressure
regulator
directly
connected to
the pneumatic
booste.
To operate, the
booster requires
an air pressure
between 3 - 6
bar
Pneumatic booster
to load pressure
into the ultraWAVE
reactor
Microwave
unit
Not
included
in the
ultraWAVE
3
package.
Not included in
the ultraWAVE 3
package
P/N SL0050
Included in
the
ultraWAVE 3
package
Air compressor
and pressure
regulator of air
are not included
in the
ultraWAVE 3
package
P/N GSB00100
Optional, can be
ordered separately
Installation of the Pneumatic booster (optional)
Booster can be placed closed to the ultraWAVE 3
unit

Table of Contents