Milestone Srl MA245-003 – ultraWAVE 3 – User Manual
52
Touch the UP button to open the cover.
When the cover stops, touch the UP button for a second
time to complete the opening.
The procedure can be interrupted touching STOP.
Once the pressure vessel is opened, the procedure stops
automatically.
Pre-conditioning of ultraWAVE 3 before the use
It is strongly recommended to run a pre-conditioning cycle on the unit before starting using it with your
samples.
This is an easy way to quickly clean the reactor and various rack components.
For pre-conditioning we suggest to carry out the method “cleaning method for vial” that you find in the
Milestone software library.
Cleaning method for pre-conditioning of the
unit