126
Run (Excel): Excel file of the final
temperature profile. This file includes all details
about the run including the microwave energy
emitted moment by moment, the values of other
sensors installed and any errors.
The Audit Trail
An audit trail (also called audit log) is a chronological record that provide documentary evidence of the
sequence of activities related to the access, profiles, settings and methods modifications.
It is the user's responsibility to verify that the software time and date are correct.
It is the user's responsibility to synchronize the software clock according to local time
also in case of solar-legal time change.
The audit trail records all the following actions (the
trail records the old and the new values):
All login events
Error messages
Changes of settings (in the Setting panel)
New users profiles
LAN settings
Rotor configurations
Sensor values
Data and time
Changing of Methods
Saving
Modifications
Deleting
Saving of the final run file
Usage of the internal
memory
Date, time and
description of the action
Description of modification
(Old and new values are
recorded)
Search Filter
Save or delete* the Trail