Milestone Srl MA245-001 ultraWAVE 3 User Manual
153
PFPE grease with
anticorrosion
additive
In case of use of
aggressive acid media:
High volume of HCl
Mixtures of HCl and
H2O2
Mixture of HCl and
H2SO4
Aqua Regia solution
(3 parts of HCl with
1 part of HNO3)
Use it on the stainless-
steel reactor, clamps
and cover
70723A
(Can be ordered
separately)
The protective oil included with the ultraWAVE 3 is silicon based, which could not be appropriate in case
of Si determination; for this reason Milestone also provides PFPE grease.
If the application requires routine use of HCl, Aqua regia or other aggressive acid media, it is
recommended to use PFPE grease with anti-corrosion additive.
For details on purchasing oil or grease for your ultraWAVE 3, contact your local Milestone supplier.
Cleaning before long operational breaks
Whenever the system is not going to be used for long period (≥ 5 days), Milestone suggests to perform a
fast cleaning cycle. The following procedure allows you to remove any acid deposits from the high pressure
lines.
Place 200 ml of deionized water with 5 ml of HNO
3
conc. into the PTFE vessel (do not use the rack) and
run the following heating program.
Load pressure: 40 bar
Step
Time
T1
T2
P
Power
1
00:15:00
200°C
60°C
100
1500W
2
00:05:00
200°C
60°C
100
1500W
At the end of the program, open the reactor and clean it. Open the reactor and remove the water, then
clean the system (with Ethanol and oil or grease) following the steps described above.