9 MAINTENANCE AND CLEANING
PROCEDURES
To secure a good and safe operation of the system, it is imperative that the
following instructions for maintenance and service are strictly followed.
The maintenance intervals can vary according to the use / application of the system and when
used under extremely aggressive operating conditions (chemicals, temperature).
Optimal maintenance not only extends the life of the system but reduces consumables costs.
The non-observation of the service and maintenance instructions contained herein,
it will void any warranty.
In the event of problems that can be corrected only by Milestone Service Engineers,
immediately stop operating the UltraWAVE 3 system and contact your local representative.
For cleaning and service always use safety garments (glasses, gloves, lab coat and in general
protective clothing).
Cleaning after each run
Cleaning of reactor and checking of the main PTFE vessel
Frequency:
Every run
Dry the reactor with a tissue and clean it with a dry paper moistened.
Do not use hard and/or sharp-edged objects/tools for cleaning but only soft clothes. The gap between
pressure vessel and PTFE bottom plate can be cleaned with a brush.
Cleaning of pressure vessel