Milestone Srl MA245-001 – ultraWAVE 3 – User Manual
75
Vessel Covering
At the End of the working day, before the unit switching off
Leave the PTFE vessel in the reactor when the unit is off. This is to protect the reactor from
any dust or condensation.
In case there is high humidity in the laboratory, it is recommended to put some absorbent paper on the
reactor
The clamps are carried in the position
represented in this picture. Pay attention that
the pressure vessel and the locking bolt are
not in direct contact each other, otherwise
there will be no drying process.
Place absorbent paper sheet on the clamps.
The paper cannot completely seal the pressure
vessel, so that the remaining moisture can
leak from the vessel.
Leak Test of the main reactor
In case of pressure decrease after the heating phase (when the temperature is
constant) of 0.5 bar every 10 min.
After preventive maintenance by Milestone service engineer
Leakage of the ultraWAVE 3 is tested with the following protocol. If the leak test is not
passed, it is necessary to check the installation of all parts and components like PTFE
cover and O-rings as well as PTFE-TFM rings. If the system is still leaky, consult
Milestone or local representative service.
Create the following program (for operating instructions see the Software Manual)