Use of authority checks to ensure
that only authorized individuals can
use the system, electronically sign a
record, access the operation or
computer system input or output
device, alter a record, or perform
the operation at hand.
The access to the system is divided in 5 levels
- Supervisor
- Service
- Administrator
- Manager
- User
The Supervisor and Service levels are reserved to
Milestone organization personnel.
The rights of Administrator, Manager and User levels are
specified on the User Manual.
The access to the software can be further secured by the
use of individual and electronically signed USB sticks.
At total, it is possible to create up to 16 login accounts
with Administrator, Manager and User rights.
The software does not allow to set simultaneously two
accounts with the same name and it is possible to define
an expiration time for the passwords.
It is not possible to use the same password immediately
after the expiration of the previous.
Password must have a minimum length of 5 digits and
must be modified at the first access of a new account.
Failed login attempts are registered in the Trail and after
3 consecutive unsuccessful attempts from the same
account, this is blocked. The release of a blocked account
can only be authorized from the following login from a
superior level or automatically after 24 hours.
Use of device (e.g., terminal)
checks to determine, as
appropriate, the validity of the
source of data input or operational
instruction.
The microwave system with terminal and dedicated
easyCONTROL software are a "Closed System". The input
of data on the terminal from the user can be done via
touch screen, keyboard or mouse.
The software recognizes the validity of the entered data
and immediately notifies the user in case of invalidity or
the same.
There is not any other external data source except this
means.