The following table are just guidelines. If it is necessary to digest larger quantities of samples
than those listed below, contact the Milestone specialist.
LOW REACTIVITY
Soils, aqueous solutions, feed, biological high-water content
ultraWAVE 3 racks
27 pos
20 pos
7 pos
Sample amount per vials
up to 0.5 g
up to 1 g
up to 3 g
Total sample amount
20 g
20 g
21 g
MEDIUM REACTIVITY
Food with low fat content, pharma, API, biological
ultraWAVE 3 racks
27 pos
20 pos
7 pos
Sample amount per vials
up to 0.25 g
up to 0.5g
up to 1.5 g
Total sample amount
10 g
10 g
10.5 g
HIGH REACTIVITY
Polymer, petrochemical, oils, food high fat content, solvents
ultraWAVE 3 racks
27 pos
20 pos
7 pos
Sample amount per vials
up to 0.1 g
up to 0.25 g
up to 0.8 g
Total sample amount
4 g
5 g
5.6 g