7 Settings Related to Operation
100
7.7 Setting the F key operation
Any of the following operations can be assigned to . To assign an operation, set the connected
number to function F-55.
Operation Number
No assignment
0
Display value hold
1
Zero set
2
Zero set clear
3
Print command through serial interface
4
Transition to COMP setting mode
5
Confirmation of CC-Link station number and RS-422/485 ID
6
Forced termination of multimode
7
Described below are the F key operation setting procedures.
1. Press once to display .
2. Press .
The Function mode is activated, and the display changes to a two-digit numerical value prefixed
by [F-].
3. Using and , change the numerical value to , then press .
The number showing the current F key operation appears.
0 (No assignment)
For setting procedures with and , see Section 1.6, "Basic key operation".
4. Using , set the F key operation.
Whenever is pressed, the number is cyclically changed from 0 to 7 in ascending order.
5. Press .
6. Press twice to return to Measurement mode.
Default
[F-55]