EasyManuals Logo
Home>Minebea>Measuring Instruments>CSD-709

Minebea CSD-709 User Manual

Default Icon
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #130 background imageLoading...
Page #130 background image
10 Communications via USB
128
10.4 Remote calibration via USB
In Command mode, you can remotely calibrate the CSD-709 by specifying any of the following
numbers in addition to command number 90 (Execute the calibration).
No. Operation relative to calibration
01
Transition to calibration mode
02
Setting unit weight
03
Setting measuring weight
04
Setting weight mass
07
Zero calibration (actual load based)
08
Zero calibration (numerical value input based)
10
Span calibration (actual load based)
11
Span calibration (numerical value input based)
21
Setting decimal point display position
40
Confirmation of calibration status
99
Registration
00
Forced termination

Table of Contents

Other manuals for Minebea CSD-709

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea CSD-709 and is the answer not in the manual?

Minebea CSD-709 Specifications

General IconGeneral
BrandMinebea
ModelCSD-709
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals