10 Communications via USB
128
10.4 Remote calibration via USB
In Command mode, you can remotely calibrate the CSD-709 by specifying any of the following
numbers in addition to command number 90 (Execute the calibration).
No. Operation relative to calibration
01
Transition to calibration mode
02
Setting unit weight
03
Setting measuring weight
04
Setting weight mass
07
Zero calibration (actual load based)
08
Zero calibration (numerical value input based)
10
Span calibration (actual load based)
11
Span calibration (numerical value input based)
21
Setting decimal point display position
40
Confirmation of calibration status
99
Registration
00
Forced termination