10 Communications via USB
129
Execute the remote calibration using the following steps. During remote calibration, is
observed in the load window.
CAUTION
Usually, you should execute the zero calibration (actual load based) using
calibration number 07.
To execute span calibration (actual load based) using calibration number 08, you
should prepare a weight of at least two-thirds of the measuring weight.
For more information on actual load based and numerical value input based calibrations, see Section 3.1,
"Overview".
Transition to calibration mode
using calibration number 01
Setting measuring weight
using calibration number 03
Setting decimal point display
position using calibration number 21
Registration using
calibration number 99
Span calibration (actual load based)
using calibration number 10
Setting unit weight using
calibration number 02
Zero calibration (actual load based)
using calibration number 07
Common steps between actual load based and numerical value input based calibrations
Actual load based
calibration
Span calibration (numerical value input
based) using calibration number 11
Zero calibration (numerical value input
based) using calibration number 08
Numerical value input
based calibration
Setting weight mass using
calibration number 04
Confirmation of calibration status
using calibration number 40
Confirmation of calibration status
using calibration number 40

Table of Contents

Other manuals for Minebea CSD-709

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea CSD-709 and is the answer not in the manual?

Minebea CSD-709 Specifications

General IconGeneral
BrandMinebea
ModelCSD-709
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals