10 Communications via USB
121
Command Command number
Execute the calibration 90
Execute the stream transfer 99
CAUTION
If an error occurs during the data transfer in the Command mode, the data will be
returned in the following format.
This time, the Error code field has the numerical value describing the error
generated, among those listed below.
Numerical value
Error description
01 The operation cannot be executed because the command is
received in a mode other than Measurement mode.
02 Zero set error
10 Parity error
11 Stop bit detection error
12 Read error
13 An error occurs in the data after the command number.
14 The specified command number does not exist.
Referencing the load data via command numbers 20 to 26
When you specify command number 20 (Reference the displayed load), 21 (Reference the track), or
22 (Reference the peak), the following data format is returned: