7 Settings Related to Operation
89
7.2 Setting the display update frequency
To update the frequency of the value displayed in the load window, use function F-04. Described
below are the display update frequency setting procedures.
1. Press once to display .
2. Press .
The Function mode is activated, and the display changes to a two-digit numerical value prefixed
by [F-].
3. Using and , change the numerical value to , then press .
The number showing the currently set display update frequency appears.
Number Update frequency
0 4 times/s
1 20 times/s
2 50 times/s
3 100 times/s
1 (20 times/s)
For setting procedures with and , see Section 1.6, "Basic key operation".
4. Using , set the update frequency.
Whenever is pressed, the number is cyclically changed from 0 to 3 in ascending order.
5. Press .
6. Press twice to return to Measurement mode.
Default
[F-04]