9 Troubleshooting
111
Error code Meaning
Behavior of CSD-709
Remedial action
Displayed value is
smaller than the lower
limit of the allowable
range.
The error code is
displayed.
Enter a value that stays within the
allowable range.
Result of AD conversion
is overflown.
The error code is
displayed.
Enter a value that stays within the
allowable range.
Result of AD conversion
is overflown.
The error code is
displayed.
Enter a value that stays within the
allowable range.
Actual load used in the
calibration exceeds the
specified value.
The error code
flashes for two
seconds.
After the error code flashes for two
seconds, it is reset automatically.
However, when the simplified zero
calibration is executed and the span
therefore deviates from the specified
range, even returning to Measurement
mode will not resume measurement, and
the error code will continue to flash. In this
case, execute span calibration to make
the span fall within the specified range.
Actual load used in
calibration is smaller
than the specified value.
The error code
flashes.
After the error code flashes for two
seconds, it is reset automatically.
Zero point used in
calibration exceeds the
specified value.
The error code
flashes.
After the error code flashes for two
seconds, it is reset automatically.
Zero point used in
calibration is smaller
than the specified value.
The error code
flashes.
After the error code flashes for two
seconds, it is reset automatically.