E Voltage Output Card CSD-709-29
161
8. Press .
The current load appears. Adjust the input from the measuring unit so that it will be the
maximum display value.
9. Using and , adjust the voltage of the maximum display value, then press .
appears.
10. Press .
The calibration data are registered and the VCAL mode is completed.
11. Press twice to return to Measurement mode.