EasyManuals Logo

Minebea CSD-709 User Manual

Default Icon
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #48 background imageLoading...
Page #48 background image
3 Calibration
46
1. Press once to display .
2. Press three times to change the display to .
3. Press to change the display to .
4. Press once to change the display to .
5. Press .
The CADJ mode is activated and appears.
POINT
To terminate the fine adjustment, press to leave CADJ mode. When you leave CADJ mode
during the process, all settings established beforehand will be permanently erased.
6. Press .
The current zero-point load flashes.
7. Using and , set the zero-point load to [0].
Holding down decreases the value continuously. Similarly, holding down increases
the value continuously.
CAUTION
Be careful that load fluctuation induced by vibration, for example, will not be exerted
on the measuring unit during calibration.
When the input from the measuring unit exceeds 2.5 mV/V, flashing
appears. When the input is lower than -2.5 mV/V, flashing appears.
These outliers are outside the scope of zero point calibration.
8. Press .
appears.
9. Place the weight lower than the set measuring weight to the measuring unit and press .
The input from the measuring unit flashes.
10. Using and , set the displayed value to the weight mass.
CAUTION
When the input from the measuring unit exceeds 3.1 mV/V, flashing
appears. When the input is lower than the value used in the zero point adjustment,
flashing appears. These outliers are outside the scope of span
calibration.
11. Press .
appears.

Table of Contents

Other manuals for Minebea CSD-709

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea CSD-709 and is the answer not in the manual?

Minebea CSD-709 Specifications

General IconGeneral
BrandMinebea
ModelCSD-709
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals