Chapter 3 SETTINGS FOR FUNCTIONS
Chapter 3
Setting the Output Conditions
3-29
ZW
User’s Manual
Auto peak [AUTOPK] Holds the "maximum value" of the measurement result. The output
changes every time the maximum value is updated.
Auto bottom [AUTOBT] Holds the "minimum value" of the measurement result. The output
changes every time the minimum value is updated.
Auto peak to peak
[AUTOPP]
Holds the difference between the maximum and minimum values
of the measurement result. The output changes every time the
maximum or the minimum value is updated.
Average [AVE] Holds the average measured value during the sampling period.
The output changes at the end of the sampling period and is held
until the end of the next sampling period.
Sample [SAMPLE] Holds the measured value at the start of the sampling period. The
output changes at the start of the sampling period and is held until
the start of the next sampling period.
Setting [Display] Description
Current
measured value
Output
(Max.value)
Current
measured value
Output
(Min.value)
0
Current
measured value
Output
(Max.value - Min.value)
Current
measured value
Output
(Average)
Sampling period
Current
measured value
Output
Sampling period

Table of Contents

Other manuals for Omron ZW Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron ZW Series and is the answer not in the manual?

Omron ZW Series Specifications

General IconGeneral
BrandOmron
ModelZW Series
CategoryAccessories
LanguageEnglish

Related product manuals