3-28
Chapter 3 SETTINGS FOR FUNCTIONS
ZW
User’s Manual
Chapter 3
Setting the Output Conditions
Setting the Hold
Set the hold conditions for measured values. The hold functions hold any value from the
measured values during the specific period (sampling period), such as the maximum or
minimum value.
■ Type
Set the hold conditions for measured values.
Setting [Display] Description
Peak [PEAK] Holds the maximum value during the sampling period. The output
changes at the end of the sampling period and is held until the end
of the next sampling period.
Bottom [BOTTOM] Holds the minimum value during the sampling period. The output
changes at the end of the sampling period and is held until the end
of the next sampling period.
Peak to Peak [P-P] Holds the difference between the maximum and minimum values
during the sampling period. This option is selected mainly when
detecting vibration. The output changes at the end of the sampling
period and is held until the end of the next sampling period.
Current
measured value
Output
Max.value
Sampling period
Current
measured value
Output
Min.value
Sampling period
Current
measured value
Output
(Max.value - Min.value)
Max.value
Sampling period
Min.value