Chapter 4 I/O SETTINGS
Chapter 4
Settings for I/O
4-23
ZW
User’s Manual
■ Output format
●
When the measured value is output
Outputs the measured value as a 21-bit binary number (2's complement).
(Example1)
When the measured value is "+1.23456" (The number of digits after the decimal point is
the default (5 digits).)
123456 (decimal) 000011110001001000000 (binary)
(Example2)
When the measured value is "-1.23456"
-123456 (decimal) 111100001110111000000 (binary)
When multiple tasks are output continuously, the currently output task number is output in the
BINARY_OUT output.
Checking Information p.3-47
6 Press or keys to select either of "TASK1"
and press key.
7 Enable TASK1 binary output.
The current setting value is displayed on the sub-
display.
Press / keys to enter the editing mode, and the
sub-display blinks.
Select "ON" and press key.
8 Press key for two seconds to enter the RUN
mode.
Steps
Key operation Display Description
ZERORST/ZERORST/
ESCESC
ZERO/ZERO/
SETSET
ZERORST/ZERORST/
ESCESC
ZERO/ZERO/
SETSET
RUN
FUN
TEACH