About Installation and Connection
When Sensor Head calibration fails
If the Sensor Head is calibrated in an inappropriate environment, an error is displayed on the main
display. If this happens, press key to return to the previous screen and try again. If an error
continues even after calibrating the Sensor Head in an appropriate environment, the fiber connector on
the Sensor Head or controller may be stained. Clean the fiber cable referring to p.1-16.
5 Press key for two seconds to enter the RUN
mode.
Steps
Key operation Display Description
RUN
FUN
TEACH