Chapter E: CALIBRATIONS 1 ADJUSTING AND CALIBRATING X-RAY BEAM
Technical manual Planmeca ProOne 99
Observe the position of the patient support table from behind the tube head. The line on
the underside of the alignment ruler must coincide with the y-line on the ball phantom.
ProO_Y_Line.eps
Alignment ruler
Lines must
Ball phantom
Tube head side
Sensor head side
Patient support table
coincide
C-arm in Y-position;
bottom view