EasyManua.ls Logo

R&S FSPN - Page 344

R&S FSPN
555 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSPN
344User Manual 1179.4363.02 ─ 01
Parameters:
<Voltage> <numeric value>
Range: -10 to 28
*RST: 0
Default unit: V
Example: //Define a minimum voltage of -5 V
SOUR:VOLT:CONT:LEV:LIM:LOW -5
Manual operation: See "Defining the output level" on page 125
SOURce:VOLTage:CONTrol<i>:LEVel[:STATe]
<State>
This command turns the tuning voltage source (V
tune
) on and off.
Note that DC power is actually supplied only if you additionally activate the outputs in
general.
Prerequisites for this command
Turn on DC power sources (SOURce:VOLTage[:STATe]).
Suffix:
<i>
.
1..n
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Example: //Turn on the V
tune
voltage source
SOUR:VOLT:CONT:LEV ON
Manual operation: See "Turning the output on and off" on page 124
SOURce:VOLTage:POWer<i>:LEVel:AMPLitude <Voltage/Current>
This command defines the output voltage for the V
supply
source.
Suffix:
<i>
.
1..n
irrelevant
Common Measurement Settings

Table of Contents