Remote Control
R&S
®
FSPN
363User Manual 1179.4363.02 ─ 01
[SENSe:]LIST:RANGe<r>:FREQuency:STOP?
This command queries the stop frequency offset of a half decade.
Suffix:
<r>
.
1..n
Half decade
Return values:
<Frequency> <numeric value>
Default unit: Hz
Example: //Query stop frequency offset for fifth half decade
LIST:RANG5:FREQ:STOP?
would return, e.g.
300000
Usage: Query only
Manual operation: See "Working with the half decade configuration table"
on page 115
[SENSe:]LIST:RANGe<r>:XCOunt <Count>
This command defines the number of cross-correlation operations done in a half dec-
ade.
The application of the command depends on the half decade configuration mode
([SENSe:]SWEep:MODE):
●
Automatic mode: The number of cross-correlations can only be queried, because it
is calculated based on an RBW factor.
●
Manual mode: The number of cross-correlations can be defined directly as an
absolute value.
Suffix:
<r>
.
1..n
Half decade
Parameters:
<Count> <numeric value> (integer only)
*RST: Automatic configuration mode is on.
Default unit: ---
Example: //Define 50 cross-correlation operations for the fifth half decade
SWE:MODE MAN
LIST:RANG5:XCO 50
Manual operation: See "Defining cross-correlation parameters" on page 113
See "Working with the half decade configuration table"
on page 115