Remote Control
R&S
®
FSPN
365User Manual 1179.4363.02 ─ 01
Example: //Sweep in forward direction
SWE:FORW ON
Manual operation: See "Selecting the sweep direction" on page 115
[SENSe:]SWEep:FSEGment
This command stops a measurement in any given half decade and resumes the mea-
surement of the next half decade.
Usage: Event
Manual operation: See "Defining cross-correlation parameters" on page 113
[SENSe:]SWEep:MODE <Mode>
This command selects the configuration mode of the half decade table.
Parameters:
<Mode> MANual
Manual mode: allows you to select a custom resolution band-
width and number of cross-correlations for each half decade.
Define the RBW for a half decade with [SENSe:]LIST:
RANGe<r>:BWIDth[:RESolution].
Define the number of cross-correlations for a half decade with
[SENSe:]LIST:RANGe<r>:XCOunt.
NORMal
Automatic mode: the application selects the resolution band-
width and number of cross-correlations based on the RBW and
XCORR factors.
Define the RBW factor with [SENSe:]LIST:BWIDth[:
RESolution]:RATio.
Define the XCORR factor with [SENSe:]SWEep:XFACtor.
*RST:
NORMal
Example: //Automatically select RBW and XCORR based on an RBW fac-
tor of 5 % and an XCORR factor of 1.5.
SWE:MODE NORM
LIST:BWID:RAT 5
LIST:RANG5:XCO 1.5
Manual operation: See "Defining the resolution bandwidth" on page 112
See "Defining cross-correlation parameters" on page 113
See "Working with the half decade configuration table"
on page 115
[SENSe:]SWEep:TIME <Time>
This command queries the estimated measurement time.