Remote Control
R&S
®
FSPN
361User Manual 1179.4363.02 ─ 01
11.6.6.1 Noise Configuration
Remote commands to configure the measurement range described elsewhere:
[SENSe:]FREQuency:STARt on page 359
[SENSe:]FREQuency:STOP on page 359
DISPlay:XGINdicator[:STATe]..........................................................................................361
[SENSe:]LIST:BWIDth[:RESolution]:RATio........................................................................361
[SENSe:]LIST:RANGe<r>:BWIDth[:RESolution].................................................................362
[SENSe:]LIST:RANGe<r>:FREQuency:STARt?.................................................................362
[SENSe:]LIST:RANGe<r>:FREQuency:STOP?..................................................................363
[SENSe:]LIST:RANGe<r>:XCOunt................................................................................... 363
[SENSe:]LIST:BWIDth[:RESolution]:USMall...................................................................... 364
[SENSe:]SWEep:COUNt.................................................................................................364
[SENSe:]SWEep:FORWard.............................................................................................364
[SENSe:]SWEep:FSEGment...........................................................................................365
[SENSe:]SWEep:MODE................................................................................................. 365
[SENSe:]SWEep:TIME................................................................................................... 365
[SENSe:]SWEep:XFACtor...............................................................................................366
[SENSe:]SWEep:XOPTimize[:STATe]...............................................................................366
[SENSe:]SWEep:XOPTimize:THReshold..........................................................................366
[SENSe:]TIME:STARt.....................................................................................................367
[SENSe:]TIME:STOP......................................................................................................367
DISPlay:XGINdicator[:STATe] <State>
This command turns the cross-correlation gain indicator (gray area in the Noise Spec-
trum result display) on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON
Example: //Remove the gray area from the diagram area.
DISP:XGIN OFF
Manual operation: See "Turning the display of the cross-correlation gain indicator
on and off" on page 114
[SENSe:]LIST:BWIDth[:RESolution]:RATio <Percentage>
This command defines the factor that the resolution bandwidth applied to each half
decade is based on.
More information
Parameters:
<Percentage> <numeric value>
Range: 0.1 to 30
*RST: 10
Default unit: PCT