Remote Control
R&S
®
FSPN
373User Manual 1179.4363.02 ─ 01
Suffix:
<n>
.
irrelevant
<s> irrelevant
<t> irrelevant
Return values:
<Frequency>
List of offset frequencies, one for each 10
x
spot noise position.
The number of return values depends on the measurement
range.
Default unit: Hz
Example: //Query spot noise information
CALC:SNO:DEC:X?
would return, e.g.
1000,10000,100000,1000000
Usage: Query only
Manual operation: See "Spot Noise" on page 91
CALCulate<n>:SNOise<s>[:TRACe<t>]:DECades:Y?
This command queries the vertical position of the 10
x
offset frequency spot noise infor-
mation.
Suffix:
<n>
.
Window
<s> irrelevant
<t> Trace
Return values:
<Level>
List of phase noise level values, one for each 10
x
spot noise
positions.
The number of return values depends on the measurement
range.
Default unit: dBc/Hz
Example: //Query spot noise information
CALC:SNO:DEC:Y?
would return, e.g.
-152.560974121094,-136.443389892578,
-145.932891845703,-152.560974121094
Usage: Query only
Manual operation: See "Spot Noise" on page 91
CALCulate<n>:SNOise<s>[:TRACe<t>]:DECades[:STATe] <State>
This command turns the spot noise calculation on every 10
x
offset frequency on and
off.