494User Manual 1179.4363.02 ─ 01
SYSTem:DISPlay:LOCK <State>
Defines whether the "Display Update" function remains available in remote operation
or not.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
The function remains available.
ON | 1
The function is not available and the display is not updated dur-
ing remote operation.
*RST: 0
SYSTem:DISPlay:UPDate <State>
This command turns the display during remote operation on and off.
If on, the R&S FSPN updates the diagrams, traces and display fields only.
The best performance is obtained if the display is off during remote control operation.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYST:DISP:UPD ON
Manual operation: See "Remote Display Update" on page 284
SYSTem:ERRor:CLEar:REMote
This command deletes all contents of the "Remote Errors" table.
Note: The remote error list is automatically cleared when the R&S FSPN is shut down.
Example:
SYST:ERR:CLE:REM
Manual operation: See "Display Remote Errors" on page 285
See "Clear Error List" on page 289
SYSTem:ERRor:DISPlay <State>
This command the error display during remote operation on and off.
If activated, the R&S FSPN displays a message box at the bottom of the screen that
contains the most recent type of error and the command that caused the error.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYST:ERR:DISP ON