Remote Control
R&S
®
FSPN
495User Manual 1179.4363.02 ─ 01
Manual operation: See "Display Remote Errors" on page 285
SYSTem:IDENtify:FACTory
This command resets the query to *IDN? to its default value.
Manual operation: See "Reset to Factory String" on page 283
SYSTem:IDENtify[:STRing] <String>
This command defines the response to *IDN?.
Parameters:
<String> String containing the description of the instrument.
Manual operation: See "Identification String" on page 283
SYSTem:KLOCk <State>
This command activates the local lockout (remote control) or returns to the local mode.
Parameters:
<State> ON
LLO (local lockout)
OFF
GTL (go to local)
*RST: OFF
Example:
SYST:KLOC ON
Activates LLO (remote control)
Manual operation: See "Local" on page 289
SYSTem:LANGuage <Language>
This command selects the system language.
Example:
SYST:LANG 'PSA'
Emulates the PSA.
Manual operation: See "Language" on page 286
SYSTem:LXI:LANReset
This command resets the LAN configuration, as well as the "LAN" password and
instrument description.
Manual operation:
See " LAN Reset" on page 288