Remote Control
R&S
®
FSPN
498User Manual 1179.4363.02 ─ 01
DIAGnostic:HUMS:DELete:ALL
This command deletes the complete HUMS data. This includes device history, device
tags, SCPI connections, utilization history and utilizations.
Example: //Delete HUMS data
DIAG:HUMS:DEL:ALL
Usage: Event
DIAGnostic:HUMS:FORMat <DataFormat>
This command defines the format for the queried HUMS data. You can query the
HUMS data either in JSON format or XML format.
The defined format affects all other commands that return block data.
Parameters:
<DataFormat> JSON | XML
JSON
Returns the HUMS data in JSON format.
XML
Returns the HUMS data in XML format.
*RST: JSON
Example: //Return data in JSON format
DIAG:HUMS:FORM JSON
DIAGnostic:HUMS:STATe <State>
This command turns the HUMS service and data collection on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON
Example: //Turn on HUMS service
DIAG:HUMS:STAT ON
DIAGnostic:HUMS:TAGS:ALL?
This command returns all key-value tags that you have assigend to the instrument.
Depending on the set data format, the queried data is either displayed in XML or JSON
format. For more information about setting the data format, see DIAGnostic:HUMS:
FORMat on page 498.
Return values:
<ID> ID number of the defined tag.
<Key> String containing key name of the defined tag.
<Value> String containing value of the defined tag.
Configuring the R&S FSPN