95 Performing a Bump Test GX-6000 Operator’s Manual
b. At the end of the calibration, the instrument displays the results from
both the bump test and the calibration. Use the AIR and SHIFT
(PANIC) button to scroll between the calibration/bump test results,
the bump test gas readings, and the calibration gas readings.
c. To return to the Gas Select Screen, press and release the POWER/
ENTER button at any time. If any channel failed the calibration, the
buzzer will sound a double pulsing tone until you press and release
the POWER/ENTER button.
d. Disconnect the tubing from the probe.
13 . Unscrew the demand flow regulator from the calibration cylinder.
14 . If you want to bump test any more channels, repeat Step 13 - Step 13.
15 . Use the SHIFT (PANIC) button to display the ESCAPE screen, then
press and release the POWER/ENTER button to return to the Calibration
Mode Menu.
16 . Use the SHIFT (PANIC) button to place the cursor next to the
NORMAL MODE menu option, then press and release the POWER/
ENTER button to return to Measuring Mode.
BUMP
CH4 OK OK
O2 NG NG
H2S OK OK
CO NG OK
CAL
CH4 O2
%LEL %
H2S CO
ppm ppm
50 18.0
25.0 49
BUMP
CH4 O2
%LEL %
H2S CO
ppm ppm
49 19.0
23.0 32
AIR or
AIR or
AIR or
SHIFT (PANIC)
SHIFT (PANIC)
SHIFT (PANIC)

Table of Contents

Other manuals for RKI Instruments GX-6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-6000 and is the answer not in the manual?

RKI Instruments GX-6000 Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-6000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals