162 Operator’s Manual · Version 3.0
6 Configuration Multiplate® analyzer
Electronic QC
Electronic QC
f Configuration > Electronic QC
The Electronic QC function can only be accessed with a level 2 password.
Figure 6-16 Electronic QC window
Limit for the deviation between
measurement and reference value
Enter the maximum allowed deviation between the actual electronic control
measurement and the reference value. If the deviation is higher the electronic control
measurement fails for that channel. Default value is “5%”.
Limit for the coefficient of
variation of measured raw values
Enter the maximally allowed coefficient of variation of the measured raw values. If
the CV is higher the electronic control measurement fails for that channel. Default
value is “2%”.