Roche Diagnostics
168 Operator’s Manual · Version 3.0
7 Maintenance Multiplate® analyzer
Cleaning and disinfecting the analyzer
Cleaning and disinfecting the analyzer
Surfaces, which may come into contact with potentially hazardous materials, must be
disinfected according to local regulatory standards.
p To clean the synthetic surface of the analyzer
1
Use a cloth moistened with mild detergent.
NOTICE
Damage to the analyzer due to use of inappropriate cleaning materials
Cleaning materials that contain solvents can damage the surface of the analyzer.
r Do not use detergents that contain solvents.
s
p To clean the measurement block
1
Use one of the recommended detergents
Recommended detergents for disinfecting the measurement block:
o Terralin® Liquid
o Mikrozid® Liquid
o Mikrozid® AF
o Mikrozid® tissues
o Sagrosept® tissues
s
p To clean apertures of the measurement block
1
Use cotton swabs moistened with one of the recommended detergents.
s