Roche Diagnostics
Operator’s Manual · Version 3.0 95
Multiplate® analyzer 4 Software
Menu options
LIS configuration
f Configuration > LIS configuration
Use this function to configure the LIS communication.
Figure 4-37 LIS configuration window
u LIS configuration (p. 161)
Electronic QC
f Configuration > Electronic QC
Use this function to configure the criteria for the electronic control measurements.
The Electronic QC function can only be accessed with a level 2 password.
Figure 4-38 Electronic QC window
u Electronic QC (p. 95)
Configure data alarms
f Configuration >Configure data alarms
This function is designed for configuring the data alarm parameters.
q
Because the data alarm parameters affect quality control, Roche recommends not to
change any of them.
The precision and correlation of duplicate sensor measurements are monitored
electronically. The data alarm function monitors the correlation coefficient of the two
single sensor measurements (CC), and the deviation from the mean (DIF).